Syringe Filter

Guide to choosing correct Syringe Filters

Syringe filters look simple: a small disc that screws onto a syringe. But choosing the wrong one can ruin a sample, clog an HPLC column or leave you with bacteria in a solution you thought was sterile. Every syringe filter is defined by four choices: membrane material, pore size, diameter and sterile or non-sterile. This guide walks through each one.

Quick answer

  • Aqueous samples, general lab use: nylon or hydrophilic PVDF.
  • Protein or biological samples: hydrophilic PVDF (low protein binding).
  • Strong organic solvents or gases: PTFE (hydrophobic).
  • HPLC sample clarification: 0.45 µm for most work; 0.22 µm for UHPLC and very fine columns.
  • Removing bacteria: 0.22 µm, and the filter must be sterile.
  • Volume: 13 mm for small samples (up to about 10 mL); 25 mm for about 10–100 mL.

1. Membrane material

The membrane decides which liquids the filter can handle and how much of your sample sticks to it.

MembraneNatureBest forAvoid with
NylonHydrophilicAqueous solutions, many common organic solvents, HPLC sample prepStrong acids; protein samples (high binding)
PVDF (hydrophilic)Hydrophilic, low protein bindingBiological and protein samples, HPLC, buffers, mobile phasesVery strong bases and some aggressive solvents (check compatibility)
PTFEHydrophobicAggressive organic solvents, air and gas ventingPurely aqueous samples, unless pre-wetted with alcohol
MCE (mixed cellulose ester)HydrophilicAqueous solutions, microbiology, water testingOrganic solvents

If you are unsure, check the solvent against the manufacturer’s chemical compatibility chart before filtering. A membrane that swells or dissolves can contaminate your sample with extractables.

2. Pore size: 0.22 µm or 0.45 µm?

0.45 µm is the standard choice for clarifying samples and removing particles before analysis. It flows faster and clogs less, and it suits most conventional HPLC columns.

0.22 µm (often written 0.2 µm) retains most bacteria. Use it for sterile filtration of media, buffers and heat-sensitive reagents, and for samples going onto UHPLC or sub-2 µm particle columns, where even fine particles can block the frit.

For very dirty or viscous samples, a prefilter (for example a glass-fibre layer, or a 0.45 µm filter before a 0.22 µm filter) prevents early clogging.

3. Diameter: 13 mm or 25 mm?

Diameter decides how much liquid you can filter before the membrane clogs, and how much sample stays inside the filter (the hold-up volume).

4. Sterile or non-sterile?

This is where mistakes are most costly. A 0.22 µm pore size only gives a sterile filtrate if the filter itself is sterile and individually packed, and you work aseptically.

Non-sterile filters are perfectly suited to analytical work: HPLC and GC sample preparation, clarifying reagents, and particle removal before UV-Vis or other instruments. They are also more economical in bulk packs.

The Axiva syringe filters we currently stock are non-sterile, designed for analytical sample preparation. If you need sterile filters for cell culture or microbiology, contact us and we will help you source the right product.

Our syringe filter range

ProductMembranePoreDiameter
Axiva syringe filterNylon0.22 µm13 mm
Axiva syringe filterNylon0.45 µm13 mm
Axiva syringe filterNylon0.22 µm25 mm
Axiva syringe filterNylon0.45 µm25 mm
Axiva syringe filterPVDF0.22 µm13 mm
Axiva syringe filterPVDF0.45 µm13 mm
Axiva syringe filterPVDF0.22 µm25 mm
Axiva syringe filterPVDF0.45 µm25 mm

See the full syringe filters category.

How to use a syringe filter correctly

  1. Draw the sample into the syringe first, then attach the filter. Drawing liquid back through the filter can damage the membrane.
  2. Twist the filter onto the Luer-lock tip firmly, without over-tightening.
  3. Push with steady, gentle pressure. If it suddenly gets hard to push, the membrane is clogged; change the filter instead of forcing it, which can rupture the membrane.
  4. For analytical work, discard the first few drops to rinse out any extractables.
  5. Use each filter once. Syringe filters are single-use.

Frequently asked questions

Can I use a nylon filter for HPLC?

Yes. Nylon is widely used for HPLC sample preparation with aqueous and many organic mobile phases. For protein or peptide samples, choose PVDF to reduce sample loss.

Does a 0.22 µm filter make a solution sterile?

Only if the filter is supplied sterile, the receiving container is sterile and you work aseptically. A non-sterile 0.22 µm filter removes particles, but it does not guarantee sterility.

Why is my syringe filter so hard to push?

The sample may be viscous or particle-heavy, or you may be using a 0.22 µm filter where a 0.45 µm one or a prefilter would work better. A 25 mm filter also flows more easily than a 13 mm one.

Can I filter organic solvents through PVDF?

Hydrophilic PVDF is compatible with many common organic solvents, but not all. Check the manufacturer’s compatibility chart for aggressive solvents, or use PTFE.