Guide to choosing correct Syringe Filters
Syringe filters look simple: a small disc that screws onto a syringe. But choosing the wrong one can ruin a sample, clog an HPLC column or leave you with bacteria in a solution you thought was sterile. Every syringe filter is defined by four choices: membrane material, pore size, diameter and sterile or non-sterile. This guide walks through each one.
Quick answer
- Aqueous samples, general lab use: nylon or hydrophilic PVDF.
- Protein or biological samples: hydrophilic PVDF (low protein binding).
- Strong organic solvents or gases: PTFE (hydrophobic).
- HPLC sample clarification: 0.45 µm for most work; 0.22 µm for UHPLC and very fine columns.
- Removing bacteria: 0.22 µm, and the filter must be sterile.
- Volume: 13 mm for small samples (up to about 10 mL); 25 mm for about 10–100 mL.
1. Membrane material
The membrane decides which liquids the filter can handle and how much of your sample sticks to it.
| Membrane | Nature | Best for | Avoid with |
|---|---|---|---|
| Nylon | Hydrophilic | Aqueous solutions, many common organic solvents, HPLC sample prep | Strong acids; protein samples (high binding) |
| PVDF (hydrophilic) | Hydrophilic, low protein binding | Biological and protein samples, HPLC, buffers, mobile phases | Very strong bases and some aggressive solvents (check compatibility) |
| PTFE | Hydrophobic | Aggressive organic solvents, air and gas venting | Purely aqueous samples, unless pre-wetted with alcohol |
| MCE (mixed cellulose ester) | Hydrophilic | Aqueous solutions, microbiology, water testing | Organic solvents |
If you are unsure, check the solvent against the manufacturer’s chemical compatibility chart before filtering. A membrane that swells or dissolves can contaminate your sample with extractables.
2. Pore size: 0.22 µm or 0.45 µm?
0.45 µm is the standard choice for clarifying samples and removing particles before analysis. It flows faster and clogs less, and it suits most conventional HPLC columns.
0.22 µm (often written 0.2 µm) retains most bacteria. Use it for sterile filtration of media, buffers and heat-sensitive reagents, and for samples going onto UHPLC or sub-2 µm particle columns, where even fine particles can block the frit.
For very dirty or viscous samples, a prefilter (for example a glass-fibre layer, or a 0.45 µm filter before a 0.22 µm filter) prevents early clogging.
3. Diameter: 13 mm or 25 mm?
Diameter decides how much liquid you can filter before the membrane clogs, and how much sample stays inside the filter (the hold-up volume).
- 13 mm: small volumes (up to about 10 mL), such as HPLC vials, with low sample loss.
- 25 mm: larger volumes (about 10–100 mL), such as buffers, media and mobile phases, with faster flow.
- Above about 100 mL: switch to a vacuum filtration setup with 47 mm membrane discs, such as Millipore HAWP MCE membranes (0.45 µm) or Millipore Durapore PVDF membranes (0.45 µm).
4. Sterile or non-sterile?
This is where mistakes are most costly. A 0.22 µm pore size only gives a sterile filtrate if the filter itself is sterile and individually packed, and you work aseptically.
Non-sterile filters are perfectly suited to analytical work: HPLC and GC sample preparation, clarifying reagents, and particle removal before UV-Vis or other instruments. They are also more economical in bulk packs.
The Axiva syringe filters we currently stock are non-sterile, designed for analytical sample preparation. If you need sterile filters for cell culture or microbiology, contact us and we will help you source the right product.
Our syringe filter range
| Product | Membrane | Pore | Diameter |
|---|---|---|---|
| Axiva syringe filter | Nylon | 0.22 µm | 13 mm |
| Axiva syringe filter | Nylon | 0.45 µm | 13 mm |
| Axiva syringe filter | Nylon | 0.22 µm | 25 mm |
| Axiva syringe filter | Nylon | 0.45 µm | 25 mm |
| Axiva syringe filter | PVDF | 0.22 µm | 13 mm |
| Axiva syringe filter | PVDF | 0.45 µm | 13 mm |
| Axiva syringe filter | PVDF | 0.22 µm | 25 mm |
| Axiva syringe filter | PVDF | 0.45 µm | 25 mm |
See the full syringe filters category.
How to use a syringe filter correctly
- Draw the sample into the syringe first, then attach the filter. Drawing liquid back through the filter can damage the membrane.
- Twist the filter onto the Luer-lock tip firmly, without over-tightening.
- Push with steady, gentle pressure. If it suddenly gets hard to push, the membrane is clogged; change the filter instead of forcing it, which can rupture the membrane.
- For analytical work, discard the first few drops to rinse out any extractables.
- Use each filter once. Syringe filters are single-use.
Frequently asked questions
Can I use a nylon filter for HPLC?
Yes. Nylon is widely used for HPLC sample preparation with aqueous and many organic mobile phases. For protein or peptide samples, choose PVDF to reduce sample loss.
Does a 0.22 µm filter make a solution sterile?
Only if the filter is supplied sterile, the receiving container is sterile and you work aseptically. A non-sterile 0.22 µm filter removes particles, but it does not guarantee sterility.
Why is my syringe filter so hard to push?
The sample may be viscous or particle-heavy, or you may be using a 0.22 µm filter where a 0.45 µm one or a prefilter would work better. A 25 mm filter also flows more easily than a 13 mm one.
Can I filter organic solvents through PVDF?
Hydrophilic PVDF is compatible with many common organic solvents, but not all. Check the manufacturer’s compatibility chart for aggressive solvents, or use PTFE.